Multiparameter Patient Monitor MD9000s Meditech

Multiparameter Patient Monitor MD9000s Meditech

Multiparameter Patient Monitor MD9000s Meditech patient monitor has a 12.1"color TFT screen. Monitoring ECG, SP02, NIBP, RESP .2-TEMP. PR as standard parameters, and 2-IBP, ETC02 and thermal printer as optional. MD9000s is one of its kind and the best selling patient monitor in Europe.

Features:

  • 12.1'' TFT color LCD, multi-language interface(Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakh, Polish, Czech).
  • Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
  • All-round monitor for adult, pediatric and neonate.
  • With standard interface, oxygen graph, trend graph, big character interface and view bed, convenient to observe. Finish all operations by keys and knobs.
  • Maximum 8-channel waveform synchronous display.
  • Display 7-lead ECG waveform on one screen, cascade ECG waveform display
  • Adopt digital SpO2 technology, anti-motion and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
  • Heart rate variability (HRV) analysis function
  • NIBP measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP data.
  • Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
  • Drug concentration calculation and titration table functions.
  • One-touch printing of trend graph
  • Connect to Central Monitoring System by 3G, Wi-Fi or wired mode.
  • AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
  • Anti-high frequency surgical unit, defibrillation-proof (special leads are necessary).
Options:
  • Invasive Blood Pressure (IBP)
  • Carbon dioxide (CO2) Module
  • Masimo SpO2 Module
  • Thermal Printer
Specification: ECG Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL, V) Lead selection I, II, III, avR, avL, avF, V, Waveform 2 ch Lead mode 3 Leads ( R, L, F or RA, LA, LL) Lead selection I, II, III, Waveform 1 ch Gain ´2.5mm/mV, ´5.0mm/mV, ´10mm/mV, ´20mm/mV HR and Alarm Range Adult 30 ~ 300 bpm Neo/Ped 30 ~ 350 bpm Accuracy ±1% or ±1bpm, which great Resolution 1 bpm Sensitivity > 200 uV P-P Differential Input Impedance > 5 MΩ CMRR Monitor > 105 dB Operation > 105 dB Diagnosis > 85 dB Electrode offset potential ±300mV Leakage Current < 10 uA Baseline Recovery < 10 s After Defi. ECG Signal Range ±8 mV (Vp-p) Bandwidth Surgery 1 ~ 15 Hz Monitor 0.5 ~ 35 Hz Diagnostic 0.05 ~ 100 Hz Calibration Signal 1 mV (Vp-p), ±5% Accuracy ST Segment Monitoring Range Measure and Alarm -2.0 ~ +2.0 mV ARR Detecting Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC Alarm Available Review Available Tall T-wave rejection capability 1.2mV Heart rate averaging the average value of the latest 6 R-R intervals which have ignored the maximum and minimum Updating rate of the display 1s Heart rate meter accuracy and response to irregular rhythm: Bigeminy ventricular 80bpm Bigeminy ventricular alternative lente 57 or 63bpm Bigeminy ventricular alternative rapid 122 or 123bpm Systoles bidirectional 95bpm Response time to heart rate meter to change in heart rate 80 to 120bpm Average 3.7s Range 3.2s ~ 4.1s 80 to 40bpm Average 4.7s NIBP Method Oscillometric Mode Manual, Auto, STAT Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120,240,480,960Min Measuring Period in STAT Mode 5 Min Pulse Rate Range 40 ~ 240 bpm Alarm Type SYS, DIA, MEAN Measuring and alarm Range Adult Mode SYS 40 ~ 270 mmHg DIA 10 ~ 215 mmHg MEAN 20 ~ 235 mmHg Pediatric Mode SYS 40 ~ 200 mmHg DIA 10 ~ 150 mmHg MEAN 20 ~ 165 mmHg Neonatal Mode SYS 40 ~ 135 mmHg DIA 10 ~ 100 mmHg MEAN 20 ~ 110 mmHg Resolution Pressure 1mmHg Accuracy Pressure Maximum Mean error ±5mmHg Maximum Standard deviation ±8mmHg Overpressure Protection Adult Mode 297±3 mmHg Pediatric Mode 240±3 mmHg Neonatal Mode 147±3 mmHg TEMPERATURE Channel 2 Measuring and Alarm Range 0 ~ 50 °C Resolution 0.1°C Accuracy ±0.1°C Actualization interval about 1 Sec. Average Time Constant < 10 Sec. IBP (optional) Channel 2 Label ART, PA, CVP, RAP, LAP, ICP, P1, P2 Measuring and alarm range ART 0~300mmHg PA -6~120mmHg CVP/RAP/LAP/ICP -10~40mmHg P1~P2 -10~300mmHg Press Sensor Sensitivity 5 uV/V/mmHg Impedance 300-3000Ω Resolution 1mmHg Accuracy ±2% or 1mmHg which great Actualization Interval about 1 Sec Display Device 12.1 in. Color TFT Made by Sharp Japan Messages 7 Waveforms Maximum 1 Alarm LED ( Yellow/Red) Screenshot 1 Power LED (Green) 1 Battery Charge LED ( Yellow) 3 Sound Mode corresponding Alarm Mode Signal Interface ECG Output BNC Amplitude 1 V/mV Accuracy < 5 % Impedance 50 Ω Signal Delay < 20 mS Recall Trend Recall Short 1 hrs, 1 Second Resolution Long 72 hrs, 1 Min. Resolution Alarm Event Recall 71 alarm events of all parameters and 8/16/32seconds of corresponding waveform. NIBP Measurement Recall At least 400 NIBP measurement data. Battery Rechargeable 3.7 A/Hr 7.4V Li battery Operating time under the normal use and full charge greater than 180minutes Operating time after the first alarm of low battery will be about 5 minutes Recorder (Option) Record Width 48 mm Paper Speed 25/50 mm/S Trace 2 Size and Weight Size Monitor 310 x 140 x 263 mm Weight Monitor 3.8 kg

Touch Screen Patient Monitor MD9015

Touch Screen Patient Monitor MD9015

Touch Screen Patient Monitor MD9015 is a touch screen 15 inch patient monitor by Meditech 192-hour trend graph,1000 group of blood pressure list,ST segment measurements, arrhythmia analysis and more functions, Support up to 32 TCP/IP network connection,composed of a central information system.. Comes with remote control. Please see our complete range of Patient Monitor & Multiparameter Monitors Features:
  • 5” high resolution color TFT display with maximum 13 waveforms
  • Full touch-screen enables intuitive operation by clicking on the specific parameters  of the waveform in real time, standard configuration includes the navigation knob
  • Pacemaker detection, electrosurgical interference proof
  • Defibrillation protection and defibrillation synchronization
  • SD Card slot enable memory extension for large data storage
  • USB and serial port enable abundant future upgrades
  • VGA output and Analog output
  • OxyCRG available to judge the respiration and circulation function for neonate
  • Large font display
  • Arrhythmia and S-T segment analysis
  • Built-in rechargeable Li-ion (6hours maximum with 2x 4200mAh batteries)
  • Nurse Call function and bi-directional communication with MFM-CMS central station
  • Multi language: English/Simplified Chinese/Traditional
  • Chinese/Spanish/Czech/Italian/Polish/French/German/Brazilian/Portugese/Russian.
  • Standard configuration: 3/5-lead ECG, NIBP, SpO2, RESP, 2-Temp, PR
  • Optional: 12-lead ECG, Nellcor-SpO2, Omron NIBP, 2-IBP/4-IBP, Mainstream/Side stream CO2,
  • CO, Multi-Gas, Full Touch Screen, Thermal Recorder, WLAN Accessory
  • Large Data Storage: 120 hours graphic and tabular trends of all parameters; 1200 sets of NIBP
  • measurement and 70 alarm events; 2-USB ports and SC card slot ensure data storage/function upgrades.
  • Clinical data can be viewed and managed via Patientcare viewer software (To be released)
  • Clinical network: bi-directional communication with CMS-2000 central station by wired or wireless connection,
  • Complete and flexible mounting solution for ICU and operation room. High-end parameters for more flexible clinical applications.
Options:
  • Invasive Blood Pressure (IBP)
  • Carbon dioxide (CO2) Module
  • Masimo SpO2 Module
  • Thermal Printer

Touch Screen Patient Monitor MD9015

Specification: Electrocardiogram (ECG)
  • Input mode Five-lead ECG input / three-lead ECG input
  • Lead selection I, II, III, aVR, aVL, aVF, V-
  • Gain selection AUTO,0.25X,0.5X,1X,2X,4X
  • Operating mode Diagnostic mode, monitoring mode, operation mode
  • Calibration signal 1mV ±5%
  • Protection The isolation voltage 4000V, 50/60Hz. Anti-high-frequency interference of electric knife.
  • Anti-defibrillation protection.
  • Pacemaker at the same time use Support
  • Electrode off instructions Each electrode (RL excluded)
  • Scanning speed 12.5mm/s, 25 mm/s, 50 mm/s
  • Common mode rejection ratio ≥90dB
  • Input impedance ≥5.0MΩ
Heartrate (HR)
  • Measuring range Adult 30 ~ 300bpm, newborns 30 to 350 bpm
  • Resolution 1 bpm
  • Accuracy < ±5bpm
  • Heart rate sensitivity ≥0.20mVp-p
  • Alarm range 15 to 350 bpm, upper and lower limits
  • Alarm indication Audible and visual alarm, display the data and parameters flashing display alarm message text prompts, there are three levels of alarm
Pulse Oximetry
  • SPO2 Range: 1% to 99%
  • Resolution: 1%
  • Accuracy: 70% to 99% range: ±2%;
  • 1% to 69% range: ±5%;
  • Method: Dual wavelength LED
ST-segment
  • ST-segment measurement range -2.0~+2.0mV
  • ST-segment measurement accuracy Within the range of -0.8 ~ +0.8 mV: ± 0.02mV or reading± 10%, Whichever is greater; Other range: undefined
  • Resolution 0.01mV
  • ST-segment update cycle 10s
  • ST segment alarm range -2.0 ~ +2.0 MV, the upper and lower limits continuously adjustable
Arrhythmia analysis
  • Arrhythmia types 13 kinds of (scalable)
  • Respiration Rate
  • Rate: 2 to 120 bpm (CO2)
Resolution: 1 breath/min NIBP (Non-invasive Blood Pressure)
  • Technique: Oscillometric during inflation
  • Range: Adult/Pediatric: 40 to 240 mmHg
  • Neonate: 10 to 150 mmHg
  • Measurement Cycle: < 40 sec. typical
Automatic Measurement
  • Cycles (Selectable): 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
  • STAT Mode: 5 minutes of continuous readings
  • Max. Allowable Cuff
  • Pressure: Adult: 300mmHg
  • Pediatric: 240 mmHg;
  • Neonate: 150 mmHg
  • Resolution: 1 mmHg
  • Transducer Accuracy: ±2 mmHg or ±2% of reading,
  • Heart (Pulse) Rate
  • Range: 30 to 250 bpm
  • Accuracy: ±2 bpm or ±3%, whichever is greater
EtCO2 technological index
  • Interface form External interface Range of
  • measurement 0~10 kPa or (0~75) mmHg
  • precision(error) (0kPa~5.4kPa) the error is ±0.3kPa、
  • (5.5kPa~10.0kPa)the error is ±0.7kPa
  • Alarm upper limit 2.1~13.2 kPa
  • Alarm lower limit 0~6.5 kPa
  • time 12s
  • EtCO2 exist
  • Unit kPa / mmHg / %
  • Compatibility
  • external main stream and side stream are compatible
IBP technological index Measured channel Single way and double way are optional Range of measurement SYS: 6.7kPa~32.0kPa(50mmHg~240mmHg) MEAN: 3.4kPa~26.6kPa(26mmHg~200mmHg) DIA: 2.9kPa~24.0kPa(22mmHg~180mmHg) Mean error ≤±0.4kPa (±3mmHg) Range of alarm SYS: upper limit 0.3kPa~53.3kPa(2mmHg~400mmHg) Lower limit 0kPa~53.1kPa(0mmHg~398mmHg) MEAN: upper limit 0.3kPa~53.3kPa (2mmHg~400mmHg) Lower limit 0kPa~53.1kPa (0mmHg~398mmHg) DIA: upper limit 0.3kPa~53.3kPa ( 2mmHg~400mmHg) Lower limit 0kPa~53.1kPa ( 0mmHg~398mmHg) Measured unit mmHg/kPa Specification of sensor DPT-248(space) Consumables mode SCW-UTAH-01