Patient Monitor MD9012 Meditech
Patient Monitor MD9012 Meditech
Patient Monitor MD9012 Meditech is 12 inch TFT display high resolution. Automatic temperature drift treatment of NIBP measuring patented technology;Friendly and configurable user interface;Powerful data review:Maximum 720-hour graphic and tabular trends, Up to10000 pieces of NIBP record storage, Wired or wireless networkable with Central Monitoring System;On-line upgrading keeps you updated.
Features:
- High resolution 12” color TFT display
- 12 Channel waves.9 Parameters : ECG, Resp, NIBP, SpO2, TEMP,
- Light weight, compact and portable
- Built-in rechargeable lithium battery
- Unique human voice alarm
- Rich screen layouts: Standard, Big font, Trend table, NIBP trend, ECG full lead, OxyCRG, IoC Layout
- Powerful storage: Maximum 720 Hour graphic and tabular trends of all
- parameters, up to 10000 NIBP review, 200 alarm review ,2 Hours of all wave forms
- System software can be upgraded through network
- Data and waveforms color be adjustable
- Arrhythmia analysis and S-T segment analysis
- Suitable for adult, pediatric and neonatal patient
Options:
- Invasive Blood Pressure (IBP)
- Carbon dioxide (CO2) Module
- Thermal Printer
- Touch Screen
Specification:
ECG
Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V)
Lead selection I, II, III, avR, avL, avF, V,
Waveform 2 ch
Lead mode 3 Leads ( R, L, F or RA, LA, LL)
Lead selection I, II, III,
Waveform 1 ch
Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto
HR and Alarm
Range
Adult 15 ~ 300 bpm
Neo/Ped 15 ~ 350 bpm
Accuracy ± 1% or ± 1bpm,which great
Resolution 1 bpm
Sensitivity > 200 (uV )
Differential Input Impedance > 5 MΩ
CMRR
Monitor > 105 dB
Operation > 105 dB
Diagnosis > 85 dB
Electrode offset potential ±300mV
Leakage Current < 10 uA
Baseline Recovery < 3 S After Defi.
ECG Signal Range ±8 m V ( Vp-p )
Bandwidth
Surgery 1 ~ 15 Hz
Monitor 0.5 ~ 35 Hz
Diagnostic 0.05 ~ 100 Hz
Calibration Signal 1 (mV ), Accuracy :±5%
ST Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY,MISSED BEATS, PNP, PNC
Alarm Available
Review Available
RESPIRATION
Method Impedance between R-F(RA-LL)
Differential Input Impedance >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
Base line Impedance Range: 0 – 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range
Adult 0 ~ 120 rpm
Neo/Ped 0 ~ 150 rpm
Resolution 1 rpm
Accuracy ±2 rpm
Apean Alarm 10 ~ 40 S
NIBP
Method Oscillometric
Mode Manual, Auto, STAT
Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
Measuring Period in STAT Mode 5 Min
Pulse Rate Range 40 ~ 240 bpm
Alarm Type SYS, DIA, MEAN
Measuring and alarm range
Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
MEAN 20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
MEAN 20 ~ 165 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution
Pressure 1mmHg
Accuracy
Pressure
Maximum Mean error ±5mmHg
Maximum Standard deviation ±8mmHg
Overpressure Protection
Adult Mode 297±3 mmHg
Pediatric Mode 240±3 mmHg
Neonatal Mode 147±3 mmHg
SpO2
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy 70% ~ 100% ±2 %
0% ~ 69% unspecified
Actualization interval about 1 Sec.
Alarm Delay 10 Sec.
Pulse Rate
Measuring and Alarm Range 0~254bpm
Resolution 1bpm
Accuracy ±2bpm
TEMPERATURE
Channel 1
Measuring and Alarm Range 0 ~ 50 C
Resolution 0.1~C
Accuracy ±0.1~C
Actualization interval about 1 Sec.
Average Time Constant < 10 Sec.
IBP
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
ART 0 ~ 300 mmHg
PA -6 ~ 120 mmHg
CVP/RAP/LAP/ICP -10 ~ 40 mmHg
P1/P2 -10 ~ 300 mmHg
Press Sensor
Sensitivity 5 uV/V/mmHg
Impedance 300-3000Ω
Resolution 1 mmHg
Accuracy ±2% or ±1mmHg, which great about 1 Sec.
Patient Monitor MD9012 Meditech
Patient Monitor MD9012 Meditech is 12 inch TFT display high resolution. Automatic temperature drift treatment of NIBP measuring patented technology;Friendly and configurable user interface;Powerful data review:Maximum 720-hour graphic and tabular trends, Up to10000 pieces of NIBP record storage, Wired or wireless networkable with Central Monitoring System;On-line upgrading keeps you updated.
Features:
- High resolution 12” color TFT display
- 12 Channel waves.9 Parameters : ECG, Resp, NIBP, SpO2, TEMP,
- Light weight, compact and portable
- Built-in rechargeable lithium battery
- Unique human voice alarm
- Rich screen layouts: Standard, Big font, Trend table, NIBP trend, ECG full lead, OxyCRG, IoC Layout
- Powerful storage: Maximum 720 Hour graphic and tabular trends of all
- parameters, up to 10000 NIBP review, 200 alarm review ,2 Hours of all wave forms
- System software can be upgraded through network
- Data and waveforms color be adjustable
- Arrhythmia analysis and S-T segment analysis
- Suitable for adult, pediatric and neonatal patient
Options:
- Invasive Blood Pressure (IBP)
- Carbon dioxide (CO2) Module
- Thermal Printer
- Touch Screen
Specification:
ECG
Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V)
Lead selection I, II, III, avR, avL, avF, V,
Waveform 2 ch
Lead mode 3 Leads ( R, L, F or RA, LA, LL)
Lead selection I, II, III,
Waveform 1 ch
Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto
HR and Alarm
Range
Adult 15 ~ 300 bpm
Neo/Ped 15 ~ 350 bpm
Accuracy ± 1% or ± 1bpm,which great
Resolution 1 bpm
Sensitivity > 200 (uV )
Differential Input Impedance > 5 MΩ
CMRR
Monitor > 105 dB
Operation > 105 dB
Diagnosis > 85 dB
Electrode offset potential ±300mV
Leakage Current < 10 uA
Baseline Recovery < 3 S After Defi.
ECG Signal Range ±8 m V ( Vp-p )
Bandwidth
Surgery 1 ~ 15 Hz
Monitor 0.5 ~ 35 Hz
Diagnostic 0.05 ~ 100 Hz
Calibration Signal 1 (mV ), Accuracy :±5%
ST Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY,MISSED BEATS, PNP, PNC
Alarm Available
Review Available
RESPIRATION
Method Impedance between R-F(RA-LL)
Differential Input Impedance >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
Base line Impedance Range: 0 – 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range
Adult 0 ~ 120 rpm
Neo/Ped 0 ~ 150 rpm
Resolution 1 rpm
Accuracy ±2 rpm
Apean Alarm 10 ~ 40 S
NIBP
Method Oscillometric
Mode Manual, Auto, STAT
Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
Measuring Period in STAT Mode 5 Min
Pulse Rate Range 40 ~ 240 bpm
Alarm Type SYS, DIA, MEAN
Measuring and alarm range
Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
MEAN 20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
MEAN 20 ~ 165 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution
Pressure 1mmHg
Accuracy
Pressure
Maximum Mean error ±5mmHg
Maximum Standard deviation ±8mmHg
Overpressure Protection
Adult Mode 297±3 mmHg
Pediatric Mode 240±3 mmHg
Neonatal Mode 147±3 mmHg
SpO2
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy 70% ~ 100% ±2 %
0% ~ 69% unspecified
Actualization interval about 1 Sec.
Alarm Delay 10 Sec.
Pulse Rate
Measuring and Alarm Range 0~254bpm
Resolution 1bpm
Accuracy ±2bpm
TEMPERATURE
Channel 1
Measuring and Alarm Range 0 ~ 50 C
Resolution 0.1~C
Accuracy ±0.1~C
Actualization interval about 1 Sec.
Average Time Constant < 10 Sec.
IBP
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
ART 0 ~ 300 mmHg
PA -6 ~ 120 mmHg
CVP/RAP/LAP/ICP -10 ~ 40 mmHg
P1/P2 -10 ~ 300 mmHg
Press Sensor
Sensitivity 5 uV/V/mmHg
Impedance 300-3000Ω
Resolution 1 mmHg
Accuracy ±2% or ±1mmHg, which great about 1 Sec.
You must be logged in to post a review.
MAECENAS IACULIS
Vestibulum curae torquent diam diam commodo parturient penatibus nunc dui adipiscing convallis bulum parturient suspendisse parturient a.Parturient in parturient scelerisque nibh lectus quam a natoque adipiscing a vestibulum hendrerit et pharetra fames nunc natoque dui.
ADIPISCING CONVALLIS BULUM
- Vestibulum penatibus nunc dui adipiscing convallis bulum parturient suspendisse.
- Abitur parturient praesent lectus quam a natoque adipiscing a vestibulum hendre.
- Diam parturient dictumst parturient scelerisque nibh lectus.
Scelerisque adipiscing bibendum sem vestibulum et in a a a purus lectus faucibus lobortis tincidunt purus lectus nisl class eros.Condimentum a et ullamcorper dictumst mus et tristique elementum nam inceptos hac parturient scelerisque vestibulum amet elit ut volutpat.
Related products
Automated Hematology Analyzer DC 2400 Plus
Automated Hematology Analyzer DC 2400 Plus
Automated Hematology Analyzer DC 2400 Plus is one of the best Fully automated hematology analyzer which can perform two duties at the same time monitoring of the blood cells having diseased cells and at the same time counting them. DiA CELL DC-2400 PLUS is very useful machine for the patients, laboratory staff etc, some of the features are: Features:- Perfect QA
- Simple to Operate
- Convenient Maintenance
- Advanced Design
Carry Over: RBC, WBC, HGB <0.5%, PLT<1.0%
Operating Interface: Colorful Windows interface, no need connecting computer. Operating by graphic button with mouse and keyboard
Display: Big size 10.4’’ TFT LCD, all parameters and histograms displayed in the same screen
Print Mode: Internal high-speed thermal printer. Optional connection with specified external printer. English printouts
Reagents System: Open reagent system. No cyanogen reagents. Provide original reagents.
Connection Port: USB, RS-232, network interface
Power: Maximum 150VA
Environment: Temperature 10℃-35℃. Relative humidity: 10%-90%
Power Supply: 220(1±10%)V, (50±1)Hz. 90V-260V, 50-60Hz broad range power available for different Countries
Dimension: 520mm×460mm×420mm
Weight: 23Kg
BP Aneroid Riester Germany
BP Aneroid Riester Germany
BP Aneroid Riester Germany isn't your average aneroid blood pressure monitor. Sure, it fits snugly in the palm of your hand, making it portable and convenient. But within that compact design lies a powerhouse of features that cater to both healthcare professionals and self-monitoring individuals.Patient Monitor Accuit Sign 8 MDX-USA
Patient Monitor Accuit Sign 8 MDX-USA
Patient Monitor Accuit Sign 8 MDX-USA is one of the best patient monitor equivalent to any other renowned brand, MDX USA has made it with sophisticated technology which makes Patient Monitor Accuit Sign 8 MDX-USA standing top quality patient monitor. please click here to download the product catalog Accuit Sign™ 8.pdf Feature:- High resolution 12.1"color TFT display
- Suitable for adult, pediatric and neonatal patient
- Color TFT Screen, waveform up to 8 channels
- Operation modes: Diagnostic, Monitor, Surgery
- Light and portable with Built-in rechargeable lithium battery
- UP to 10 kinds of monitor parameters
- Arrhythmia analysis, pace-marker detection and S-T Segment analysis
- 72 hour storage and review of trend grams and tables
- 500 NIBP measurement date can be storage and recall
- Electro surgical unit and defibrillation protected
- Networkable with central monitoring system
- 5-lead or 3-lead selectable
- Input: 5-lead :RA; LA; RL; LL; V or R; L; N; F; C
- ECG waveforms of 7-leads display on the same screen
- Lead selection :I; II; III; avR; avL; avF; Vx
- ECG waveform :2 channels
- Gain selection :12.5mm/s, 25mm/s, 50mm/s
- Accuracy :±1bpm or ±1%, whichever is greater
- Protection :Withstand 4000VAC/50Hz voltage in isolation against electrosurgical and defibrillation
- Calibration signal :1mV±5%
- Alarm range: Adult :15-300bpm
- Pediatric/Neonatal :15-350bpm
- Arrhythmia analysis and S-T segment analysis
- S-T segment detection Measurement rang :-0.2mv~2.0mv
- Alarm :Yes, audible and visual alarm, alarm events review
- Operation modes :Manual/Automatic/STAT
- Measurement unit :mmHg/KPa selectable
- Measurement types :Systolic, Diastolic and Mean
- Over-pressure protection :YES
- Resolution :1mmHg
- Accuracy :Mean error and standard deviation per ANSI/AAMI SP-10
- Alarm :Systolic, Diastolic, and Mean
- Auto measure time :Adjustable
- Measurement rang :10-270mmHg
- 500 NIBP measurement date can be storage and recall
- Measurement range :0-100%
- Accuracy :70-100%, ±2% 0-69%, upspecified
- Alarm range :0-100% Pulse rate
- Range :20-254bpm
- Accuracy :±3bpm
- Respiration
- Method :RA-LL impedance
- Pediatric/Neonatal :6-150rpm
- Alarm range: Adult :6-120rpm
- Pediatric/Neonatal :6-120rpm
- Apnea alarm :YES
- Channel :Dual-channel
- Alarm range :0-50℃
- Apnea :0.1℃